While running the HAPSITE SMART on the Service Module, I received an ‘SM over temp’ error. What can I do to correct this?

Applies to: Chromatography Smart (GC/MS) | Tickets Solved: 0 | Rank: 3.5

Error: SM over temp

The over temp error is common in a Service Module whose oil wick and diaphragm have become deficient. Power off the Service Module and HAPSITE and allow the power to drain from both instruments. Wait approximately one hour before turning on the instruments and attempting to complete the previous operation.

This is a temporary fix; to address this issue permanently, call technical support.