An error message appears while trying to calibrate the IdentiFINDER, "Cannot adjust fine gain. Remove all sources and press AGAIN." How can I correct this error?

Applies to: Radiation Identifinder (Gamma Spectroscopy) | Tickets Solved: 0 | Rank: 4.5

This particular error while calibrating the IdentiFINDER can be caused by radiation sources in the area, a weak calibration source, or by an extreme change in temperature. To overcome this error remove all sources or radiation and/or try a different cailbration source. If the area is particularly hot or cold wait for the IdentiFINDER to adjust to the external temperatures. For further questions please contact technical support.